search Icon
×
Positioning iconCurrent location:HOME>NEWS>Technical Articles>Online Particle Counter (HM-CLZ1) Product Knowledge Graph White Paper

Online Particle Counter (HM-CLZ1) Product Knowledge Graph White Paper

Article Source:Hengmei Cleanliness Instrument Network  Release time:2025-12-03 17:50:23

1. Introduction

1.1 Document Version

This white paper is Version V1.0.

1.2 Compilation Unit

Shandong Hengmei Electronic Technology Co., Ltd.

Address: Oulong Science and Technology Park, No. 1 Jinma Road, Weifang City, Shandong Province

1.3 Compilation Date

2025.12.3

1.4 Core Positioning

This white paper presents the full-dimensional product information of the Hengmei HM-CLZ1 Online Particle Counter in a structured. It covers technical parameters, functional features, application scenarios, and operational guidelines to serve as a standardized reference for R&D, production, sales, and maintenance personnel.

2. Table of Contents

1. Introduction

2. Table of Contents

3. Product Basic Information

4. Core Technologies and Functional Features

5. Detailed Technical Parameters

6. Application Scenarios and Industry Adaptation

7. Accessories and Consumables Configuration

3. Product Basic Information

3.1 Product Name

Hengmei Intelligent Online Particle Counter

3.2 Product Model

HM-CLZ1 (Online Type)

3.3 Product Description

The HM-CLZ1 Online Particle Counter automatically determines cleanroom grades in accordance with the confidence measurement and verification standards specified in the national standard GB/T6167-2007 and calibration specification JJF1190-2008. It can simultaneously measure particle counts across six particle size channels in a single sampling cycle and display the number of particles in each channel along with their real-time changes. Equipped with a 7-inch touchscreen and Android 7.1 operating system, it supports data storage, network transmission, and multi-standard cleanroom grade 判定,making it suitable for measuring the size and quantity of suspended particles in unit volume of gas.

3.4 Reference Standards

• National Standard: GB/T6167-2007 Performance Test Methods for Particle Counters

• Cleanliness Grade Standards: ISO14644, FS209E, New GMP (Dynamic/Static)

• Calibration Specification: JJF1190-2008

4. Core Technologies and Functional Features

4.1 Core Technologies

• Light Scattering Sensing Technology: The core light scattering sensor significantly eliminates stray light in the cavity, greatly improving the signal-to-noise ratio for more accurate and reliable measurements. It uses an imported semiconductor laser diode (service life >30,000 hours) to ensure long-term stable operation.

• Multi-Parameter Sensing Technology: Built-in temperature, humidity, and atmospheric pressure sensors to collect environmental parameters, providing contextual references for measurement accuracy.

• Data Processing & Transmission Technology: Calculates counting results based on 95% confidence level (UCL) and converts them into m³ and ft³ units. Supports multiple data transmission methods for efficient data storage and sharing.

4.2 Functional Features

1. Multi-Parameter Monitoring: Integrates temperature, humidity, and atmospheric pressure sensors to monitor environmental conditions alongside particle counts.

2. Audible & Visual Alarm: Triggers audible and visual alarms for particle counts exceeding user-set thresholds via configurable alarm parameters.

3. Enhanced User Experience: Runs on Android 7.1 operating system with a 7-inch color touchscreen, enabling intuitive and smooth operation.

4. Automatic Grade Determination: Automatically judges cleanroom grades based on cleanroom classification standards, simplifying operation.

5. Versatile Data Transmission: Supports WiFi, Ethernet (wired) for cloud platform uploads, and USB flash drive for data copying (driver-free plug-and-play).

6. Cloud Management Support: Comes with a supervision platform client; test results are directly transmitted to the cloud for long-term/short-term cleanliness analysis and auxiliary management.

7. Flexible Sampling Modes: Offers scheduled sampling and continuous sampling to meet diverse testing needs.

8. Large-Capacity Data Storage: Stores 10,000–100,000 data records, with query and export functions for historical data tracing.

5. Detailed Technical Parameters

5.1 Host Basic Parameters

• Host Dimensions: 210mm (Length) × 150mm (Width) × 65mm (Height)

• Host Weight: 1.8 kg

• Packaging Dimensions: 350mm × 150mm × 300mm

• Packaging Weight: 2.8 kg

• Power Supply: 12V/3A (with dedicated power adapter)

• Power Consumption: 30W

• Display: 7-inch color touchscreen

• No Built-in Battery

5.2 Measurement Performance Parameters

• Particle Size Channels: 6 channels (0.3 μm, 0.5 μm, 1.0 μm, 3.0 μm, 5.0 μm, 10.0 μm)

• Sampling Flow Rate: 2.83 L/min (0.1 cfm, with automatic constant flow adjustment)

• Maximum Sampling Concentration: 35,000 particles/L (for particles ≥0.5 μm)

• Repeatability Relative Error: <10% (under consistent measurement conditions)

• Particle Size Distribution Error: ≤±30% (for 0.5 μm particles)

• Particle Concentration Indication Error: ≤±30% (for 0.5 μm particles)

• Self-Cleaning Time: ≤10 minutes

• Sampling Cycle: 10 seconds – 99 hours

• Sampling Delay: ≤60 seconds

• Measurement Interval: ≤60 seconds

• Sampling Modes: Continuous sampling, scheduled sampling

5.3 Environmental Sensing Parameters

• Temperature Measurement: Range (-40℃ to 85℃), Error (±0.3℃), Repeatability (0.1℃) (built-in sensor)

• Humidity Measurement: Range (0–100% RH), Error (±2% RH), Repeatability (0.1% RH) (built-in sensor)

• Atmospheric Pressure Measurement: Range (300–1100 hPa), Accuracy (±1 hPa), Resolution (0.1 hPa) (built-in sensor)

• Operating Environment: Temperature (0℃–40℃), Humidity (10%–75% RH); sampling gas must be free of corrosive gases (e.g., acid, alkali)

• Storage Environment: Temperature (0℃–45℃), Humidity (0–90% RH)

5.4 Data-Related Parameters

• Data Communication: Standard USB, WiFi; Optional RS485, LAN

• Data Storage Capacity: 10,000–100,000 records (queryable and exportable)

• Cleanliness Standards Supported: National Standard (GB), ISO, 209E, GMP

• Verification Standards: JJF1190-2008, GB/T6167-2007

• Power-On Auto-Start: Supports automatic sampling (per last settings) after power-on (configurable)

• Auto-Upload: Enables cloud platform uploads of test results after sampling (configurable)

6. Application Scenarios and Industry Adaptation

6.1 Application Scope

Meets the cleanroom (area) testing needs of the pharmaceutical, electronics, food hygiene, filter manufacturing, optics, and aerospace industries for precision processing, precision testing, and bioengineering.

6.2 Industry-Specific Adaptation

• Pharmaceutical Industry: Clean production workshops (pharmaceutical factories), testing laboratories (drug inspection institutes), operating rooms, and sterile wards (hospitals) – ensures compliance with clean standards for production, testing, and medical environments.

• Electronics Industry: Chip production workshops (semiconductor factories), precision machinery processing workshops – prevents particle interference in electronic component production to improve product quality.

• Food Hygiene Industry: Dairy processing plants, sealed meat product factories, condiment factories, agricultural product precision processing workshops – monitors cleanliness to avoid food contamination and ensure food safety.

• Filter Manufacturing Industry: Tests filter quality and efficiency by comparing particle concentrations before and after filtration to verify filter performance.

• Optics & Aerospace: Precision optical component processing workshops, precision test laboratories, bioengineering labs – provides clean environment monitoring for high-precision production and testing to ensure product accuracy.

7. Accessories and Consumables Configuration

7.1 Standard Accessories

• 1 HM-CLZ1 Online Particle Counter Host

• 1 12V/3A Power Adapter

• 1 Isokinetic Sampling Probe

• 1 High-Efficiency Filter

• 1 10cm Air Tube

• 2 Sampling Port Rubber Caps

• 1 Product Manual

• 1 Calibration Report

• 1 Set of Expansion Screws

7.2 Optional Consumables

• External Temperature/Humidity Sensor: For high-precision environmental monitoring (avoids temperature interference from the host; connects to the expansion interface).

• RS485/LAN Communication Module: For users requiring RS485 or LAN communication functions.

• Spare High-Efficiency Filters and Rubber Caps: For long-term maintenance.


Article address:https://www.cleanlinesstesting.com/news2/75.html