search Icon
×
Positioning iconCurrent location:HOME>PRODUCT>Infrared gas analyzer>Infrared CO/CO2 analyzer
Infrared CO/CO2 analyzer

Infrared CO/CO2 analyzer

【Introduction】Detected gases: Carbon monoxide (CO) and carbon dioxide (CO2) in air.
Measurement range: 0 to 50.0 x 10-6 CO, 0 to 0.500% CO2, or 0 to 200.0 x 10-6 C, 0.0 to 1.000% CO2.
CO/CO2 resolution: 0.1 x 10-6 CO; 0.001% CO2

  • Model iconModel: HM-Q3

Product details

I. Product Overview

The infrared CO/CO2 analyzer is a high-precision instrument for detecting toxic and hazardous substances in the air of public places. It utilizes the non-dispersive infrared (NDIR) principle to measure the concentrations of carbon monoxide and carbon dioxide in the air. It offers high accuracy, long service life, and minimal crosstalk, making it an essential instrument for environmental and occupational health monitoring.

II. Scope of Application

This instrument is widely used in environmental protection, environmental monitoring, health supervision, occupational health, disease control, and research institutes.

III. Standards 

JJG635-2011 "Carbon Monoxide and Carbon Dioxide Infrared Gas Analyzer"

HJ965-2018 "Ambient Air - Automatic Determination of Carbon Monoxide - Non-diffusive Infrared Method"

GB/T18204.2-2014 "Public Place Hygiene Inspection Methods - Part 2: Chemical Pollutants"

GB 9801-1988 "Air Quality - Determination of Carbon Monoxide - Non-diffusive Infrared Method"

GBZ/T 300.37-2017 "Determination of Air Toxic Substances in Public Places - Part 37: Carbon Monoxide and Carbon Dioxide"

IV. Technical Features 

1. Android operating system and 7-inch touchscreen display for enhanced real-time human-computer interaction.

2. Large storage capacity, capable of storing over 100,000 data records.

3. Built-in thermal printer for printing current data upon test completion.

4. Built-in Wi-Fi for uploading completed test data to the platform for convenient data monitoring (optional 4G module).

5. USB data port for transferring data to a USB flash drive. 6. Digital flow sensor, real-time flow data display.

7. High-precision infrared sensor for high measurement accuracy, fast response, and short warm-up time.

8. Supports nitrogen and catalytic zero calibration.

9. Internal constant temperature heating, suitable for outdoor operation in low-temperature environments.

10. Built-in channel switching solenoid valve for automatic switching between zeroing and measurement.

11. Temperature and humidity measurement function, with a built-in temperature and humidity compensation algorithm, eliminates the effects of temperature changes on measurement data.

12. Displays 0.5-hour, 8-hour, and daily average values; timer storage function.

V. Technical Parameters:

1. Detection Principle: Non-dispersive infrared gas analysis/Non-dispersive infrared (NDIR)

2. Detected Gases: Carbon monoxide (CO), carbon dioxide (CO2) in air

3. Detection Method: Built-in pump-type

4. Measuring Range: Option 1: 0-50.0×10-6 CO, 0-0.500% CO2 (0-50ppm CO, 0-5000ppm CO2)

Option 2: 0-200.0×10-6 CO, 0-1.000% CO2 (0-200ppm CO, 0-10000ppm CO2)

5. Operating Temperature Range: CO 0∽45°C, CO2 -20°C~65°C

6. Humidity: 10-95% RH

7. Concentration display: ppm, mg/m3, % (automatic conversion)

8. Resolution: 0.1×10-6 CO; 0.0001% CO2

9. Display: 7-inch capacitive touch screen

10. Storage capacity: >100,000 sets of measurement data

11. Automatic zero point correction technology

12. Linearity error: ≤±2% F·S

13. Repeatability: ≤1.0%

14. Span drift: ≤±2% F·S/3h

15. Zero drift: ≤±1% F·S/h

16. Carbon monoxide interference: 1250mg/m3CO ≤±0.3% F·S

17. Response time: CO: t0-t90 ≤45s; CO2: ≤60s

18. Warm-up time: 5 minutes

19. Flow Volume: (0.5-2.0) L/min

20. Output: USB for direct output to a USB flash drive

21. Sound and light alarm

22. Power supply: Built-in lithium battery, AC/DC dual-use, manually switchable

23. Dimensions: 257.3*147*300 mm